The Testo 435-3 version of Testo 435 multi-function climate measuring instrument has an integrated differential pressure sensor with a 0-25 hPa measuring range. Differential pressure measurement enables you to check filters in ventilation and air conditioning systems or to measure air velocity with a Pitot tube.
Hence Testo 435-3 multi-function climate measuring instrument with its optional probes ideally equips you for adjusting ventilation and air conditioning systems or assessing air quality.
It has attachable probes (optional) for assessing air quality, IAQ probe (Indoor Air Quality) for assessing air quality – measures CO2, air temperature, air humidity and absolute pressure.

Features
- Includes Pitot tube for differential pressure measurement for filter monitoring or for measurement
- Further optional probes can be attached: thermal and vane anemometers, along with probes for air humidity, CO2, absolute pressure and temperature
- Display of dew point distance, min., max. and mean values
- Illuminated display

Major Applications
- Differential pressure measurement on filters
- Measurement at the air outlet
- Measurement of the air and surface temperature
- Measurement of radiated heat
- Lighting

Technical Specifications
- Temperature - NTC
• Measuring range: -50 to +150 °C
- Resolution: 0.1 °C
- Temperature - TC Type K (NiCr-Ni)
• Measuring range: -200 to +1370 °C
- Accuracy: ±0.3 °C (-60 to +60 °C), ±(0.2 °C + 0.5 % of mv) (Remaining Range)
- Temperature - TC Type T (Cu-CuNi)
• Measuring range: -200 to +400 °C
- Humidity - Capacitive
• Measuring range: 0 to +100 %RH
- Absolute Pressure
• Measuring range: 0 to +2000 hPa
- Velocity - Vane
• Measuring range: 0 to +60 m/s
- Velocity - Hot wire
• Measuring range: 0 to +20 m/s
- Light
• Measuring range: 0 to +100000 Lux
- Differential pressure (internal sensor) - Piezoresistive
• Measuring range: 0 to +25 hPa
- Accuracy: ±0.02 hPa (0 to +2 hPa), ±1 % of mv (Remaining Range)

General Specifications
- Operating temperature: -20 to +50 °C
- Protection class: IP54
- Battery type: Alkali manganese, mignon, type AA
- Battery life: 200 h (typical vane measurement)
- Display ligthing: Yes
- Storage temperature: -30 to +70 °C