Technical Specifications
- Max. applied measurement voltage: ±22 V DC ±60 V DC maximum rated voltage above ground
- Resistance measurement ranges: 300 mO (max. display 310.00 mO) to 3 O (max. display 3.1000 O)
- Voltage measurement ranges: DC 20 V
- Comparator functions:
• Judgment result: Hi/IN/Lo (resistance and voltage judged independently)
• Setting: Upper and lower limit, Deviation (%) from reference value; Logical AND result: PASS/FAIL, calculates the logical AND of resistance and voltage judgment results; Result display, beeper, or external I/O output, Open-collector (35 V, 50 mA DC max.)
|
General Specifications
- Display :- 31000 full digits (resistance), 199999 full digits (voltage), LED
- Interfaces :- External I/O, RS-232C, Printer (RS-232C), GP-IB (-01 suffix models only)
-Sampling rate :- Four steps, 4 ms (Extra-FAST), 12 ms (FAST), 35 ms (Medium), 150 ms (Slow)
Power supply :- 100 to 240 V AC, 50/60 Hz, 30 VA max.
|