Technical Specifications
- DC mA Measurement: Resolution 0-24 mA; Range 0.001 mA Accuracy (% of reading + counts) 0.010% + 2 µA - Temperature coefficient: ± (0.002% of reading + 0.002% of range) /°C ( < 18°C or >28°C) - Ohms Measurement (Ohms range): 0.00 O to 4000.00 O ; Accuracy: 0.015% + 0.05 O - Ohms Source (Ohms range): 1.0 O to 4000.0 O (Excitation current from measurement device 0.1 mA to 0.5 mA) - Temprature RTD: 13types of RTD's ; Resolution: 0.1°C - Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0°C (Pt10 and Cu10), 0.6°C - Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 O~400.00 O), 0.05 mA~0.8 mA (400.0 O~1500.0 O), 0.05 mA~0.4 mA (1500.0 O ~4000.0 O), excitation current (0.25 mA for Pt1000 range). - Temperature Coefficient: ±0.05°C /°C for measure, ±0.05°C /°C (< 18°C or > 28°C) for source
|
General Specifications
- Maximum voltage applied between any terminal and earth ground or between any two terminals: 30 V - Operating temperature: -10°C to 50°C - Storage temperature: -30°C to 60°C - Operating altitude: 2,000 m - Storage altitude: 12,000 m - Vibration requirements: MIL-T-28800E, Class 2 - Drop text requirements 1 m - IP Rating: IEC 60529: IP52 - Electromagnetic environment: IEC 61326-1, Portable - Safety: IEC 61010-1, Max 30 V to earth, Pollution Degree 2
|