Fluke
Fluke 718 Pressure Calibrator
Model No :
Fluke 718
Part Code :
- 718-1G: 2547148 - 718-30US: 665075 - 718-100US: 665080 - 718-300G: 2547153

Overview

Fluke 718 offers a total pressure calibration solution for transmitters, gauges and switches. This is a compact pressure calibrator about a third of the size of comparable instruments and weighs just one kilogram (two pounds). Its easy to clean pump helps protect the pump from damage and provides the ability to service the pump without disassembly, even in the field.

Features
- Pressure source and milliamp measurement to calibrate and maintain almost any pressure device
- Integrated pump is easily cleaned when accidently exposed to fluids that reduces cost of ownership and repairs and enables servicing the pump in the field
- Error % calculation for quicker pass/fail decision making in the field
- Min/Max/Hold to capture changing measurements
- New 1 and 300 psi pressure ranges mean fewer extra tools required
- 1, 30, 100 & 300 psi ranges now available

Major Applications
- Calibration of Pressure Switch
- Calibration of P/I Transmitter
- Calibration of Pressure Transmitters
- Field Verifying Process Pressure Gauges

Technical Specifications
- mA Measure Range: 0 mA to 24 mA ; Resolution: 0.001 mA; Accuracy: 0.015% + 1 count
- Fluke 718 Series
- Measure pressure: (Internal sensor)
• Accuracy: 0.025% full scale
• Functions: Zero, min/max, hold, damp
• Pressure Range: Negative 0.85 bar to 20 bar (4 ranges)/ up to 700

General Specifications
- Maximum voltage: 30 volts (non-operating)
- Temperature: -40°C to 60°C
- Operating temperature: -10°C to 55°C
- Relative humidity: 95% (10°C to 30°C); 75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
- Operating altitude: 3,000 M max
- Shock: 1 M drop test
- Vibration: Random, 2 g, 5 Hz – 500 Hz
- Safety: CSA C22.2 No. 1010.1: 1992
- EMC: EN50082-1: 1992 and EN55022: 1994 Class B