Fluke Calibration
Fluke 3130-G2M Portable Pressure Calibrator
Model No :
Fluke 3130-G2M
Part Code :
4366073

Overview

The Fluke 3130 Portable Pneumatic Pressure Calibrator is ideal for calibrating pressure transmitters, transducers, gauges and similar devices. The 3130 contains everything you need to generate, control and measure pressure, as well as read the output of the device under test (DUT). The internal, motorized pump provides an easy,efficient alternative to time consuming, tiring hand pumps.

Features
- Measure and generate pressures from vacuum to 2 MPa (300 psi, 20 bar)
- Internal pump can generate vacuum to -80 kPa (-12 psi, -0.8 bar) or pressure to 2 MPa (300 psi, 20 bar)
- Supply pressure connection allowing the use of external gas supply up to 2 MPa (300 psi, 20 bar)
- Includes variable volume for fine adjustment of pressures
- Pressure measurement accuracy of 0.025% reading ±0.01% FS
- Electrical measurement and 24 volt supply for close looped calibrations
- Measure or generate 4 to 20 mA
- Measure 0 to 30 V DC
- Powered by internal, high capacity, NiMH battery or universal ac mains adapter
- Compatible with Fluke 700P Pressure Modules

Major Applications
- For calibrating pressure transmitters, transducers, gauges
Technical Specifications
- Pressure (internal pump) : -80 kPa to 2 MPa (-12 to 300 psi, -0.8 to 20 bar)
- Pressure (external air) : 0 to 2 MPa (0 to 300 psi, 0 to 20 bar)
• mA : 0 to 24.000 mA
• Volts : 0 to 30.000 V DC
- Instrumental Measurement Uncertainty
- Pressure : 0.025% of reading ±0.01% FS
- mA : 0.015% of reading ±0.002 mA
- Volts : 0.015% of reading ±0.002 V
- Temperature Effect (All Functions)
• No effect on accuracy on all functions from 15°C to 35°C
• Add ±0.002% F.S./°C for temps outside of 15°C to 35°C

General Specifications
- Operating temperature : -10°C to +50°C
- Storage Temperature : -20°C to +60°C
- Power requirements : 12 V DC (Universal ac adapter/charger supplied)
- Battery : Internal 3800 mAh advanced NiMH pack
Operating Time
- On full charge : Approx 50 hours (Measure only or external air; no pump)
- Using internal pump : Approx 100 calibration cycles to 300 psi